| Model Number |
MMT-551LNAS |
| Medical Device Problem Code |
Failure to Cycle (1142)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump received a compromised forced sensor system error alarm while rewinding.The insulin pump was not dropped or bumped and the drive support cap appeared normal.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test.Device received a33 alarm during the basic occlusion test due to loose or protruded drive support disk.Unable to perform prime or a33 test, excessive no delivery test and occlusion test due to a33 alarm.
|
| |
|
Search Alerts/Recalls
|