| Model Number |
MMT-551NAH |
| Medical Device Problem Code |
Priming Problem (4040)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had squirts insulin during priming.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed displacement test, rewind test, basic occlusion test, prime or a33 test, excessive no delivery test and occlusion test.No prime or fill anomaly noted.(b)(4).
|
| |
|