| Model Number |
406700 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
05/26/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reported event of a damaged sheath was confirmed.The sheath tube had been torn and detached proximal to the distal tip.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the sheath tear and detachment is consistent with damage during use.
|
| |
|
Event or Problem Description
|
|
During a procedure, a part of the sheath detached and had to be retrieved.When the sheath was attempted to be removed, the sheath had broken and a piece was left inside the femoral vein.A 10fr sheath and ptca balloon were used to retrieve the piece, requiring an additional access site.It was confirmed with fluoroscopy and angiography that no fragment remained in the patient.No further intervention was needed.The procedure was finished successfully with no adverse consequences to the patient.
|
| |
|