Model Number URF-P6 |
Device Problems
Leak/Splash (1354); Material Puncture/Hole (1504)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was received for evaluation.Evaluation of the device confirmed the reported issue of hole on the distal end of the scope.The a rubber was found leaking and a protruding metal was found on the bending section.In addition, an excessive breakage on the ig (image generator) of the device was observed.The identified parts were replaced, device was repaired.Once completed, the device was tested and passed required testing and specifications.
|
|
Event Description
|
It was reported that during preparation for use, the device was found to have hole on the distal end.The device was leaking due to the hole observed.There was no patient involvement on this report.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide the review of the device history records (dhr).Please see updated sections.The device history records for this device were reviewed and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.A probable cause of the reported issue likely due to bending tube kinked occurred during use, an excessive force was applied.As stated on the ifu and as a preventive measure, the user manual states : do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section maybe damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.Olympus will continue to monitor complaints for this device.
|
|
Search Alerts/Recalls
|