| Model Number |
406962 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/19/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
During the procedure after the puncture, the triple-hemostasis valves and sheath were not connected properly, it was not straight.Three 5fr sheaths were used to complete the procedure instead and there were no adverse consequences to the patient.Three additional access sites were required for the replacement devices due to the malfunction.Further information regarding the event is pending.
|
| |
|
Event or Problem Description
|
|
The connection between the triple valves part and the sheath was not straight.There was no detachment of the device component noted.
|
| |
|