• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION LABSYSTEM PRO; AMPLIFIER AND SIGNAL CONDITIONER, TRANSDUCER SIGNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION LABSYSTEM PRO; AMPLIFIER AND SIGNAL CONDITIONER, TRANSDUCER SIGNAL Back to Search Results
Model Number 86640
Medical Device Problem Code Power Conditioning Problem (1474)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/27/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported that due to a device malfunction the procedure had to be cancelled.A lab system pro was selected for use.There was noise on the ecg which made the procedure difficult to perform therefore the procedure had to be cancelled.
 
Additional Manufacturer Narrative
The boston scientific sales representative went onsite to checked a few things before the field service engineer arrival.The sales representative found that the customer had not connected the ecg cable correctly to the amplifier and used a long 6 meter cable.The sales rep replaced the cable with a 3 meter cable and verified the connections to the amplifier, retested and found the noise issue resolved.The ecg noise was determined to be inadvertent/unintentional interaction of the product from the bsc sales rep analysis onsite and additional available information.Furthermore, bsc rep resolved the complaint by correctly connecting the ecg to the amplifier and replacing with a shorter cable.The evidence from the product record review did not identify a potential product quality issue or new patient harm.
 
Event or Problem Description
It was reported that due to a device malfunction the procedure had to be cancelled.A lab system pro was selected for use.There was noise on the ecg which made the procedure difficult to perform therefore the procedure had to be cancelled.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LABSYSTEM PRO
Common Device Name
AMPLIFIER AND SIGNAL CONDITIONER, TRANSDUCER SIGNAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key10185240
Report Number2134265-2020-08399
Device Sequence Number3016418
Product Code DRQ
UDI-Device Identifier08714729885887
UDI-Public08714729885887
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K152693
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number86640
Device Catalogue Number86640
Device Lot Number1A909595
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/27/2020
Supplement Date Received by Manufacturer08/27/2020
Initial Report FDA Received Date06/23/2020
Supplement Report FDA Received Date09/22/2020
Patient Sequence Number1
-
-