• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE VERSACARE FRAME; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE VERSACARE FRAME; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number P3200K000574
Device Problems Delayed Alarm (1011); Wireless Communication Problem (3283)
Patient Problem No Patient Involvement (2645)
Event Date 06/12/2020
Event Type  malfunction  
Manufacturer Narrative
The hillrom technician found the side communication cable had been ran over multiple times and needed to be replaced.Per hillrom¿s health hazard evaluation: during usage the pins from the port connected to the facility wall side may become damaged (broken, bent etc.).The primary root cause for this is that the communication cable is not properly removed from the wall and secured to the bed frame before transporting the bed.When the communication cable is pulled from the wall at an angle there is a potential for the pins/connector to get damaged.This is an inherent product use case issue (potential misuse) and not a product failure to meet the specified performance requirements.Nurse call systems in accordance with ul 1069 requires that the nurse call function be supervised, meaning that if for any reason the circuit is interrupted a nurse call or trouble signal is sent to the nurse¿s station.What this means is that there is no way for a sidecom® communication cable to have a bent pin, broken wire or other issue in the nurse call circuit without the facility staff having knowledge without prior cancellation of a nurse call.The versacare user manual has the following caution/warning notice: before transporting the bed, make sure that the power cord, hoses, and other equipment are properly stowed.Failure to do so could result in equipment damage.Make sure the patient, equipment, and all lines are securely placed within the perimeter of the bed for intra-hospital transport.After transport, make sure that the nurse call system cables are properly connected.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed in march 2019.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the side communication cable to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hillrom technical support received a report from the account stating the bed exit was alarming at the bed, but was not sending an alert to the nurses' station.The bed was located in room 15 at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE FRAME
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
lori daulton
1069 state route 46 east
batesville, IN 47006
8129312280
MDR Report Key10186522
MDR Text Key196628620
Report Number1824206-2020-00278
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 06/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP3200K000574
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/12/2020
Initial Date FDA Received06/23/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/23/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-