• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH GLENOSPHERE HOLDER/ IMPA CTOR; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH GLENOSPHERE HOLDER/ IMPA CTOR; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Model Number 5901-1168
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/27/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Upon completion of investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
Primary procedure, right reverse total shoulder.It was reported that on impaction, the threaded tip of the glenosphere impactor broke in the glenosphere.Surgeon reported he had been impacting at an angle due to patient anatomy.Glenosphere was brought to the back table, the threaded portion was removed.The long gray sleeve was removed from the impactor, and the broken impactor was used to complete the case successfully with a delay of approximately 10 minutes.The rep confirmed that no further information is available.
 
Event or Problem Description
Primary procedure, right reverse total shoulder.It was reported that on impaction, the threaded tip of the glenosphere impactor broke in the glenosphere.Surgeon reported he had been impacting at an angle due to patient anatomy.Glenosphere was brought to the back table, the threaded portion was removed.The long gray sleeve was removed from the impactor, and the broken impactor was used to complete the case successfully with a delay of approximately 10 minutes.The rep confirmed that no further information is available.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device was returned for evaluation and matches the reported event.During the investigation, the device inspection revealed the following: the tip of the glenosphere holder - piston shows a clean break of the tip towards the middle of the threaded area.The surface is finely fissured and the level fracture area shows shear lips at the edge of the surface.This is specific to a ductile fracture, which is caused by an excessive force applied on the device, most probably combined to the fact that the impactor was not fully engaged/ not in alignment to the glenosphere.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLENOSPHERE HOLDER/ IMPA CTOR
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
MDR Report Key10186682
Report Number0008031020-2020-01754
Device Sequence Number886191
Product Code KWS
UDI-Device Identifier07613327101027
UDI-Public07613327101027
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K130895
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5901-1168
Device Catalogue Number5901-1168
Device Lot NumberP9S53
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2020
Initial Date Received by Manufacturer 05/27/2020
Supplement Date Received by Manufacturer08/07/2020
Initial Report FDA Received Date06/23/2020
Supplement Report FDA Received Date09/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age63 YR
-
-