| Model Number |
5901-1168 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/27/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Primary procedure, right reverse total shoulder.It was reported that on impaction, the threaded tip of the glenosphere impactor broke in the glenosphere.Surgeon reported he had been impacting at an angle due to patient anatomy.Glenosphere was brought to the back table, the threaded portion was removed.The long gray sleeve was removed from the impactor, and the broken impactor was used to complete the case successfully with a delay of approximately 10 minutes.The rep confirmed that no further information is available.
|
| |
|
Event or Problem Description
|
|
Primary procedure, right reverse total shoulder.It was reported that on impaction, the threaded tip of the glenosphere impactor broke in the glenosphere.Surgeon reported he had been impacting at an angle due to patient anatomy.Glenosphere was brought to the back table, the threaded portion was removed.The long gray sleeve was removed from the impactor, and the broken impactor was used to complete the case successfully with a delay of approximately 10 minutes.The rep confirmed that no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed, since the device was returned for evaluation and matches the reported event.During the investigation, the device inspection revealed the following: the tip of the glenosphere holder - piston shows a clean break of the tip towards the middle of the threaded area.The surface is finely fissured and the level fracture area shows shear lips at the edge of the surface.This is specific to a ductile fracture, which is caused by an excessive force applied on the device, most probably combined to the fact that the impactor was not fully engaged/ not in alignment to the glenosphere.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
|
| |
|