Catalog Number 00435001313
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 03/26/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2020-02238, 0001822565-2020-02246.Concomitant medical products: poly liner, part#: 00434906506, lot#: 64287460.Unk glenoid head.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported patient underwent a shoulder revision approximately four (4) months ago due to the poly liner disengaging from the stem.The surgeon had only replaced the poly liner and kept the originally implanted stem.Approximately one (1) month post-revision, patient underwent a second revision due to the poly and stem disengaging and joint dislocation.Both the stem and poly liner were removed and replaced.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Insufficient information provided.Unable to perform a compatibility check.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: initial op notes dated 8 jan 2020 were reviewed and no deviations were noted.Revision op notes dated 10 feb 2020 were reviewed and identified open reduction and revision of right rtsa humeral component, poly spacer.Presented with increased pain, x-ray showed a dislocated shoulder.Closed reduction attempted but unsuccessful.Upon opening, the humeral poly component was noted to be displaced anterior and off the humeral stem.X-rays were not sent for mmi review as they do not enhance the investigation.Revision op notes dated 26 mar 2020 were reviewed and identified dislocation - revision of humeral component.Patient was opening a bag of carrots and felt something give in his shoulder, xrays showed a repeat dislocation.The poly spacer was again shown to be dissociated from the humeral component.Stem and poly exchanged without complication.Dhr was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUMERAL STEM NON-POROUS 13 MM STEM DIAMETER 130 MM STEM LENGTH
Common Device Name
PROSTHESIS, EXTREMITIES
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10186922
Report Number0001822565-2020-02239
Device Sequence Number10989965
Product Code HSD
UDI-Device Identifier00889024269323
UDI-Public(01)00889024269323
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K130661
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number00435001313
Device Lot Number63754535
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/01/2020
Supplement Date Received by Manufacturer10/19/2020
Initial Report FDA Received Date06/23/2020
Supplement Report FDA Received Date10/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-