MICRO THERAPEUTICS, INC. DBA EV3 CONCERTO NYLON; DEVICE, EMBOLIZATION, VASCULAR Back to Search Results
Model Number NV-3-8-HELIX
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/27/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported the coil was being used as per the ifu, but the coil did not detach within the vessel.A new coil was used, and the procedure was successfully completed.The patient's status at the time of the report was alive-no injury.It was noted that the device and any accessory devices were prepared as indicated in the ifu.
 
Additional Manufacturer Narrative
There was no instant detacher returned with the concerto coil.The implant coil appeared to be stretched; but still attached to its pushwire.The pushwire appeared to be bent at 28.0cm from the proximal end.The ai to release wire crimps, coupler to ai crimps, and coupler to pusher tube weld are present and intact; it appeared that the mechanical detachment was not attempted.The break indicator at the manual detachment was still intact; it also appeared that the manual detachment was not attempted at this location.The coin was located against the lumen stop.The shield coil was present and intact; with no damage.The concerto coil was then successfully detached on the first attempt using an in-house instant detacher.Under the microscope, the outer jacket was then removed to gain access the coin.No damage was found with the coin.The inner diameter of the lumen stop and the inner diameter of the retainer ring were found to be visually acceptable.The lumen stop inner diameter (id) was measured to be 0.00260¿ and found to be within specification (0.00220" minimum - 0.0029" maximum).The retainer ring inner diameter (id) was measured to be 0.00460¿and found to be within specification (0.00455"+/- 0.00010").All other subassemblies appeared to be normal.No other anomalies were observed.Based on the analysis performed, the concerto coil was confirmed to have ¿non-detachment¿ issue as the returned coil was still attached to the pushwire.However, the root cause could not be determined as the returned concerto coil was successfully detached on the first attempt using an-house instant detacher.There was no mechanical or manual detachment attempt was found on the returned concerto coil.In addition, the pushwire was found to be bent; and the implant coil was found to be damaged and stretched.However, the cause for damages could not be determined.There was no non-conformance to specifications identified that lead to the non-detachment issue.Since the instant detacher was not returned; therefore, any contribution of the instant detacher to the non-detachment issue could not be assessed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONCERTO NYLON
Common Device Name
DEVICE, EMBOLIZATION, VASCULAR
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
MDR Report Key10187048
Report Number2029214-2020-00586
Device Sequence Number3016483
Product Code KRD
UDI-Device Identifier00847536013391
UDI-Public00847536013391
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K090046
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/21/2021
Device Model NumberNV-3-8-HELIX
Device Catalogue NumberNV-3-8-HELIX
Device Lot NumberA727512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2020
Initial Date Received by Manufacturer 06/20/2020
Supplement Date Received by Manufacturer09/23/2020
Initial Report FDA Received Date06/23/2020
Supplement Report FDA Received Date09/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-