BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 2.5/35; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 391235
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/04/2020
Type of Reportable Event Malfunction
Event or Problem Description
An orsiro drug-eluting stent system was chosen for treatment.During unpacking, the stent detached from the balloon when removed from the protective spiral.
 
Additional Manufacturer Narrative
The complaint instrument was not returned for analysis.Therefore no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation for the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 2.5/35
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key10188491
Report Number1028232-2020-02568
Device Sequence Number3298327
Product Code NIQ
UDI-Device Identifier07640130439368
UDI-Public07640130439368
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/19/2021
Device Model Number391235
Device Catalogue NumberSEE MODEL NO.
Device Lot Number03191357
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/23/2020
Supplement Date Received by Manufacturer08/18/2020
Initial Report FDA Received Date06/24/2020
Supplement Report FDA Received Date08/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-