• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RULER; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RULER; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71674079
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 05/29/2020
Event Type  malfunction  
Event Description
It was reported that upon inspection the threads were seem to be broken.No case involved.
 
Manufacturer Narrative
The device, intended for use in treatment, was return for evaluation.Our investigation included a visual inspection of the device and confirmed the nuts are broken on the rulers which confirmed the stated failure mode.This failure mode has been previously identified.Since the manufacturing of the complaint device, a design change was implemented to eliminate/reduce the occurrence of this failure.Only the outside of the ruler was returned.The lot number is unreadable on the device.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RULER
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10188576
MDR Text Key196224779
Report Number1020279-2020-02695
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010560759
UDI-Public03596010560759
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71674079
Device Catalogue Number71674079
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2020
Date Manufacturer Received05/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-