| Model Number |
OM-8850 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/19/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during surgery the speedstitch needle was damaged at the end and would not pass the suture.The procedure was completed with a delay of under 30 min and a backup device was available.No injury to the patient was reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H10 h3,h6: the device reported, used in treatment, was not returned for evaluation.A review of manufacturing records for the reported lot number 2047483 found non-conformances or anomalies during manufacturing process related to the reported event.A complaint history review has found related failures.A visual inspection and functional evaluation cannot be performed and customer´s complaint cannot be confirmed.A review of the product/print specifications found material discrepancy used during manufacturing.The factor related to the design or manufacture of the device that lead to the failure reported include, but are not limited to: (1) supplier related issue - failed to provide material that meet specification.The failure is being assessed for recommended actions and further evaluation regarding supplier controls are currently in progress.
|
| |
|