| Catalog Number |
0070900 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the evacuator was found to be broken when the package was open.Per additional information via email from ibc representative on 15jun2020 it was noted that the suction port was broken.
|
| |
|
Event or Problem Description
|
|
It was reported that the evacuator was found to be broken when the package was open.Per additional information via email from ibc representative on 15jun2020 it was noted that the suction port was broken.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed however the cause was unknown.Visual evaluation of the returned sample noted one opened (no original packaging), used silicone bulb evacuator.It was noted that the blue suction port was broken, with the end of the port seeming chipped off.Although the reported event was confirmed, the root cause could not be determined.A potential root cause for this failure could be defective components received from supplier.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review was not performed because labelling could not have prevented the reported failure.
|
| |
|
Search Alerts/Recalls
|