| Model Number |
TN,FC500 MPL FLOW CYTOMETER 100-240V,50/60HZ |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer (fse) was on site and observed faulty signal of the amp board in location fl3.To resolve the issue the fse replaced the tarpon amp board.Bec is filing an mdr for this event based on the fda classification of the 08 jan 2018 urgent medical device recall as a class i recall on 20 nov 2018 (recall number z-0471-2019 for fc 500; recall number z-0472-2019 for epics xl/xlmcl).Bec internal identifier - case-(b)(4).
|
| |
|
Event or Problem Description
|
|
The field service engineer (fse) observed faulty signal at the fl3 location of their fc 500 flow cytometer.There was no report of death, injury, or change to patient treatment as a result of this event.
|
| |
|
Search Alerts/Recalls
|