| Model Number |
TA9035L |
| Medical Device Problem Codes |
Failure to Fire (2610); Expulsion (2933)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
06/05/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, when closing of the open wound of the anastomosis during hemicolectomy side-to-side anastomosis, the device did not fire.The surgeon had to sew the open wound.This resulted to extended surgical time of more than 30 minutes.There was unanticipated tissue loss as a result of this problem.The same event occurred on two units of the device.The surgeon also noticed that the tip of the staples pointing outside the cartridge.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the device did not fire and the staples released prematurely from the device.The reported issue could not be confirmed.The most likely root cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are also thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Search Alerts/Recalls
|