MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION IMAGER II ANGIOGRAPHIC CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION IMAGER II ANGIOGRAPHIC CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number 38270
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 06/16/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that tip detachment occurred.An imager ii catheter was selected for use in a renal embolization procedure.The imager ii was placed into the body over a.035 non-bsc guidewire.While screening the catheter through the heart it was noted that the distal part of the catheter became detached.This ended up floating off the wire and was initially lodged in the aorta.A snare was then used in an attempt to retrieve the catheter tip, but on two occasions the snare sliced through the catheter and therefore there were then three pieces of catheter tip floating through the arteries.The snare was used to successfully retrieve one of these pieces and was removed from the right circumflex femoral artery.A long different snare was then used to attempt to snare the remaining pieces which had now travelled more distal, but without success.The main body of the catheter tip has become lodged within the right posterior tibial artery and a smaller piece has also become lodged in the right dorsalis pedis artery.The right foot was viable.After approximately three hours of attempting to remove the fragments, it was decided to leave the fragments within the patient rather than operate since the patient also has metastatic cancer.The renal embolisation was completed as was the original desired procedure for the patient using a different device.Post operatively both pt and dp pulses where and remain present and the patients foot continued to be warm to touch.No adverse effects were noted by the clinical staff within two days of the event.The patient was in stable condition.
 
Event or Problem Description
It was reported that tip detachment occurred.An imager ii catheter was selected for use in a renal embolization procedure.The imager ii was placed into the body over a.035 non-bsc guidewire.While screening the catheter through the heart it was noted that the distal part of the catheter became detached.This ended up floating off the wire and was initially lodged in the aorta.A snare was then used in an attempt to retrieve the catheter tip, but on two occasions the snare sliced through the catheter and therefore there were then three pieces of catheter tip floating through the arteries.The snare was used to successfully retrieve one of these pieces and was removed from the right circumflex femoral artery.A long different snare was then used to attempt to snare the remaining pieces which had now travelled more distal, but without success.The main body of the catheter tip has become lodged within the right posterior tibial artery and a smaller piece has also become lodged in the right dorsalis pedis artery.The right foot was viable.After approximately three hours of attempting to remove the fragments, it was decided to leave the fragments within the patient rather than operate since the patient also has metastatic cancer.The renal embolisation was completed as was the original desired procedure for the patient using a different device.Post operatively both pt and dp pulses where and remain present and the patients foot continued to be warm to touch.No adverse effects were noted by the clinical staff within two days of the event.The patient was in stable condition.
 
Additional Manufacturer Narrative
Device evaluated by mfr: device analysis was performed on the returned device.It was noticed that the device showed a separation of the tip located 2 cm from the tip edge.63 cm of the device was returned, and as the device is manufactured at 6 cm long, 2 cm was not returned.Materials analysis testing of the tip was performed and revealed that the area was brittle.A kink located 23 cm from the strain relief was noticed.During analysis and due to the extremely fragility of the device, the device cracked approximately 61 cm from the strain relief.No other damage was noticed on the devices shaft.Inspection of the remainder of the device, apart from the observed damage on the box, revealed no other damage or irregularities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMAGER II ANGIOGRAPHIC CATHETER
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10191775
Report Number2134265-2020-08495
Device Sequence Number886847
Product Code DQO
UDI-Device Identifier08714729355656
UDI-Public08714729355656
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K121694
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 08/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/12/2020
Device Model Number38270
Device Catalogue Number38270
Device Lot Number0000133414
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2020
Initial Date Received by Manufacturer 06/17/2020
Supplement Date Received by Manufacturer08/04/2020
Initial Report FDA Received Date06/24/2020
Supplement Report FDA Received Date08/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
92484513-FA
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age62 YR
-
-