| Model Number |
MMT-523RNAP |
| Medical Device Problem Codes |
Mechanical Problem (1384); Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/19/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the buttons of insulin pump were not responding and unable to rewind.The customer stated that most of the buttons were unresponsive.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
All buttons functioning properly.No button error alarm during testing.No moisture or corroded keypad found.No flatted keypad.Connector was inspected and locked on lcd board.Device passed displacement test, rewind test and self test.No rewind anomaly noted.(b)(4).
|
| |
|