• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. META-TAN LAG SCREW DRIVER; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. META-TAN LAG SCREW DRIVER; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 71631440
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 06/01/2020
Event Type  malfunction  
Event Description
It was reported the defective device was found during warehouse inspection.No case involved.
 
Manufacturer Narrative
The devices, intended for use in treatment was not returned for evaluation.Without the actual product, product evaluation could not be performed and complaint could not be confirmed.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable device that can be exposed to numerous surgeries and cleaning cycles.Probable causes that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to and after each use and cleaning.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further actions are being taken at this time.We consider this investigation closed.Credit will be issued for the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
META-TAN LAG SCREW DRIVER
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10194203
MDR Text Key196425701
Report Number1020279-2020-02732
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/25/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71631440
Device Lot Number18BSP0009
Was Device Available for Evaluation? No
Date Manufacturer Received12/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-