• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II BCS ART ISRT TRL SZ 1-2 11MM RT; PRSTHSS,KN,PTLLFMRTBL,SM-CNSTRND,CMNTD,PLYMR/MTL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II BCS ART ISRT TRL SZ 1-2 11MM RT; PRSTHSS,KN,PTLLFMRTBL,SM-CNSTRND,CMNTD,PLYMR/MTL Back to Search Results
Model Number 74035213
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/01/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the defective device was found during warehouse inspection.No case involved.
 
Additional Manufacturer Narrative
The device, intended for use in treatment was not returned for evaluation.Without the actual product, product evaluation could not be performed and complaint could not be confirmed.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A review of complaint history for the listed part revealed no prior complaints for the part/lot.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable device that can be exposed to numerous surgeries and cleaning cycles.Probable causes that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to and after each use and cleaning.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further actions are being taken at this time.We consider this investigation closed.Credit will be issued for the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II BCS ART ISRT TRL SZ 1-2 11MM RT
Common Device Name
PRSTHSS,KN,PTLLFMRTBL,SM-CNSTRND,CMNTD,PLYMR/MTL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10194312
Report Number1020279-2020-02740
Device Sequence Number11615117
Product Code JWH
UDI-Device Identifier00885556157282
UDI-Public00885556157282
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111711
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74035213
Device Catalogue Number74035213
Device Lot Number16AAP2923
Initial Date Received by Manufacturer 06/01/2020
Supplement Date Received by Manufacturer12/16/2020
Initial Report FDA Received Date06/25/2020
Supplement Report FDA Received Date12/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-