| Model Number |
5640 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Vessel Or Plaque, Device Embedded In (1204)
|
| Date of Event |
05/20/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A manufacturing record review was completed and no related nonconformances were found.No product was returned for inspection.Therefore, a returned product evaluation was not completed.The account was inquired on the procedure details.The response stated that the vessel had heavy calcium.Orbital atherectomy was performed.The procedure was able to be completed though it is unclear if the tip fragment is trapped in the calcium lesion.It is likely that the turnpike spiral was damaged during use.Based on information from the account and no returned product evaluation, the most likely root cause of the issue is undeterminable.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a peripheral pta atherectomy.It was noted that the distal tip of the turnpike spiral broke off in the mid posterior tib artery, which is heavily calcified.Snaring attempts to retrieve were unsuccessful.It is reported the patient (who is stable and doing well) will come back in attempt to remove later.No further complications were reported.
|
| |
|