• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. EMPOWR 3D KNEE; EMPOWR 3D KNEETM INS, 10L 12MM, VE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. EMPOWR 3D KNEE; EMPOWR 3D KNEETM INS, 10L 12MM, VE Back to Search Results
Model Number 341-12-710
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 05/27/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as an infection.The previous surgery and the surgery detailed in this event occurred 4.2 months apart.The healthcare professional indicated there was significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at hospital and not made available to djo surgical for examination.A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection.There were no findings during this evaluation that indicate the reported device was the source or had a direct connection with the patient's infection.No other information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - patient was infected.Insert was removed as a precaution and a new one was inserted, which was of the same thickness (12mm).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR 3D KNEE
Common Device Name
EMPOWR 3D KNEETM INS, 10L 12MM, VE
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx
MDR Report Key10195561
Report Number1644408-2020-00487
Device Sequence Number3014473
Product Code JWH
UDI-Device Identifier00888912166973
UDI-Public(01)00888912166973
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K143242
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/06/2024
Device Model Number341-12-710
Device Catalogue Number341-12-710
Device Lot Number114T1043
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/27/2020
Initial Report FDA Received Date06/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age53 YR
-
-