DATASCOPE CORP. - MAHWAH CS300; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problems No Audible Alarm (1019); Unexpected Shutdown (4019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/30/2020
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed as per company standard operating procedure since the device manufacture date is greater than one year from the event date.A getinge service territory manager (stm) was dispatched to evaluate the iabp unit.The stm was unable to reproduce the issue.No failures were shown in the logs.The stm replaced the power supply assembly as a precaution and possible source of the issue.All connections were checked.The stm completed all safety, functionality and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.
 
Event Description
It was reported that during use on a patient the cs300 intra-aortic balloon pump (iabp) was intermittently shutting off.It was also later clarified that the unit shut down three times with no alarms or warning.The power was cycled and the assist therapy was able to continue.No patient harm, serious injury or adverse event was reported.
 
Event Description
It was reported that during use on a patient the cs300 intra-aortic balloon pump (iabp) was intermittently shutting off.It was also later clarified that the unit shut down three times with no alarms or warning.The power was cycled and the assist therapy was able to continue.No patient harm, serious injury or adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key10196408
MDR Text Key196547117
Report Number2249723-2020-00981
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/25/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received06/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-