MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES 90K IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES 90K IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-01
Device Problem Electro-Static Discharge (2149)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/25/2020
Event Type  malfunction  
Event Description
The recipient is reportedly experiencing intermittency followed by static and loss of lock.External equipment was exchanged and programming adjustments were made, however, the issue did not resolve.Revision surgery is scheduled.
 
Manufacturer Narrative
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed damage at the electrode ground ring and electrode.This is believed to have occurred during revision surgery.In addition, inspection revealed a dent on the bottom cover of the device.The device passed the photographic imaging inspection.System lock could not be obtained at any spacing.The no lock condition prevented some electrical tests from being performed.The device passed some of the electrical tests performed.The internal visual inspection revealed cracked epoxy on the feedthru pins.The internal visual inspection revealed loose components.The failure of this device is attributed to excessive moisture through a gross leak at the seam weld.This ultimately caused the device to cease functioning.A corrective action has been implemented as part of the corrective and preventive action system.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key10197117
MDR Text Key196917142
Report Number3006556115-2020-00500
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup,Followup
Report Date 06/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/25/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/01/2006
Device Model NumberCI-1400-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age58 YR
-
-