| Model Number |
N/A |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Implanted date: device was not implanted.Explanted date: device was not explanted.The actual device was not returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.A review of the device history record of the product code/lot# combination was conducted with no findings.
|
| |
|
Event or Problem Description
|
|
The user facility reported that the destination devices was used during a up and over leg intervention, and the sheath kinked.The procedure was completed, and the patient's condition was stable.There was no patient injury, medical/surgical intervention required.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow up no.1 to provide the completed investigation results.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.With no return of the actual device, the exact cause of the reported event cannot be definitively determined based on the available information.
|
| |
|