• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RULER; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RULER; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71674079
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 06/02/2020
Event Type  malfunction  
Event Description
It was reported that upon inspection after the procedure the ruler was broken at the collar.No delay or additional complications were reported.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection confirmed the threaded tip is broken off the ruler and has not been returned.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RULER
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10199996
MDR Text Key196571663
Report Number1020279-2020-02784
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010560759
UDI-Public03596010560759
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 08/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71674079
Device Catalogue Number71674079
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2020
Initial Date Manufacturer Received 06/02/2020
Initial Date FDA Received06/26/2020
Supplement Dates Manufacturer Received07/14/2020
Supplement Dates FDA Received08/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-