| Model Number |
20226 |
| Medical Device Problem Code |
Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the device was not returned for analysis.However, an image depicting the express ld device stuck in a 6fr introducer sheath was received.From the image it can be observed that the balloon was not inflated.The stent was also noted to be damaged and had moved proximally on the balloon.The proximal end of the stent was bunched as it entered the light green 6fr introducer sheath.
|
| |
|
Event or Problem Description
|
|
Reportable based on investigation completed on 19jun2020.It was reported that advancing difficulties were encountered.The target lesion was located in the iliac artery.A 7.0x60x75cm express ld iliac / biliary stent was selected for use.However, the physician could not advance the stent through the sheath due to severe scarring at the entry site.Both the device and the sheath was removed and another 7fr sheath was used.The procedure was completed with a different device.No patient complications reported.However, investigation revealed stent damage and stent moved on balloon.
|
| |
|