WRIGHTS LANE SYNTHES USA PRODUCTS LLC 7.0MM TI MATRIX SCREW 50MM THREAD LENGTH; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
|
Back to Search Results |
|
| Model Number |
04.639.750 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Code Available (3191)
|
| Date of Event |
01/01/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Only event year is known.Additional device product codes: mnh, mni, kwq, kwp.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2020, the patient underwent revision due to titanium matrix screw thread length and titanium matrix top loading polyaxial head pop-on pedicle screw head becoming detached in the patient post surgery.This surgery was an l4-l5 tlif and spondylisthesis reduction.The devices were easily removed.Following the surgery the hardware was examined to find that the matrix poly head tulip collet was broken/deformed as well as the matrix pedicle screw head.Patient outcome was unknown.Concomitant devices reported: locking cap (part number unknown, lot unknown, quantity unknown), rod (part number unknown, lot unknown, quantity unknown), transverse connector (part number unknown, lot unknown, quantity unknown), ti matrix top loading polyaxial head (part number 04.632.001, lot unknown, quantity 1).This report involves one (1) 7.0mm ti matrix screw 50mm thread length.This is report 1 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6 investigation summary customer quality investigation: the complaint device was not received for investigation.The following investigation is based on the image(s) provided in the attachment(s).The image(s) were reviewed, and the complaint condition of bent/deformed could be confirmed as the top rim of the pedicle screw is damaged but damage on the very top of it should not affect the polyaxial head retention.The condition of broken (2+ pieces) could not be confirmed as the device appears to be intact.Since the device was not returned, a dimensional inspection and a functional test were not able to be performed.A manufacturing record evaluation could not be performed as the lot number could not be determined from the image provided.A definitive assignable root cause could not be determined based on the provided information.During the investigation no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: b5: additional event information updated.H3, h6: the received photos was reviewed by engineering gteam: -one collet ¿finger¿ is broken off and not present in the photo of the polyaxial heads and there is damage noted on the id of the collet and on the edges that are protruding out of the bottom of the head.-there is corresponding wear/damage on the od of the screw head.-the appearance of the parts and the broken collet are indicative of the polyaxial head not being properly seated during assembly in-situ.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional event information: additional procedure details: -the polyaxial head ¿popped off¿ within a couple of days of the initial surgery.-cory noted that the polyaxial head was attached to the pedicle screw intraoperatively.The surgeon did not use a pre-assembled polyaxial screw which is consistent with his normal process.-the surgeon typically confirms proper actuation of the head before detaching the assembly tool but it was unknown if he trialed the specific screw properly.-the surgeon used the reamer which is also part of his normal process.-the surgeon final tightened the inferior level first and then distracted.-the surgeon has been using this system for over 5 years.
|
| |
|
Search Alerts/Recalls
|
|
|