BIOMET 3I OSSEOTITE® TAPERED CERTAIN® IMPLANT 6 X 13MM; DENTAL IMPLANT Back to Search Results
Model Number XIFNT613
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Information (3190)
Date of Event 06/09/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet (b)(4).Initial reporter¿s first name is not provided / unknown.A summary investigation has been completed for lack of primary stability events that do not allege a deficiency with the implant and identified that the reported event is likely due to biological factors which have an adverse effect on implant stability or is related to surgical technique and insertion torque.Due to a wide range of external factors (non-design/ non-manufacturing related), identifying a definitive root cause is generally not possible.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
Doctor reports that they were not able to achieve primary stability with implant xifnt613 in tooth site #19.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE® TAPERED CERTAIN® IMPLANT 6 X 13MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key10200506
Report Number0001038806-2020-00888
Device Sequence Number1073507
Product Code DZE
UDI-Device Identifier00844868022077
UDI-Public(01)00844868022077(17)230410(10)2018031707(241)XIFNT613
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K130949
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 06/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/10/2023
Device Model NumberXIFNT613
Device Catalogue NumberXIFNT613
Device Lot Number2018031707
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2020
Initial Report FDA Received Date06/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age37 YR
Patient Weight75
-
-