It was reported that the device failed during procedure.It was confirmed, in an additional information, that the device broke, however no pieces fell inside the wound.All fragments were disposed.
|
|
A polarstem collar reamer guide (75102205), used in treatment, was not received for evaluation.It was reported that the device failed during the procedure.It was confirmed, in an additional information request, that the device broke, however no pieces fell inside the wound.All fragments were disposed.A visual analysis could not be conducted, as the part was not returned.The production documentation was reviewed, however there were no deviations found.The review of the complaint database showed that similar cases were reported in the past, one of which led to a corrective and preventive action.A design review was performed which demonstrated that the hook may contact the reamer if it is not fully engaged in the rasp during reaming.This is likely to lead to a fracture of the hook and the spring might become loose as reported here.It is therefore assumed that foreign material prevented the hook from fully engaging with the rasp, which led the reamer to hit and fracture the hook intraoperatively.A new design of the device has been released in order to prevent the reoccurrence of this issue.The failure mode cannot be independently confirmed as the part was not returned.The root cause is attributed to insufficient design specifications.Further investigations have been initiated to review the performance of this device.This complaint has been reopened for reporting reasons, according to the current procedure.No additional information became available.Should the device be returned, this complaint will be reassessed.Smith and nephew will monitor this version of the device for further similar issues.
|