• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. DISCOVERY ELBOW DISC HUM 4X150MM LT FLANGED C

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. DISCOVERY ELBOW DISC HUM 4X150MM LT FLANGED C Back to Search Results
Catalog Number 114914
Device Problems Loose or Intermittent Connection (1371); Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/28/2020
Event Type  Injury  
Manufacturer Narrative
The reason for this revision surgery was reported as loose stem. The previous surgery and the surgery detailed in this event occurred 1. 9 years apart. The healthcare professional indicated there was no delay in surgery and another suitable device was available for use. The revision surgery was completed as intended. The devices were disposed of at hospital and not made available to djo surgical for examination. A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements. There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event. The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery. Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review. The root cause of this complaint was a revision surgery due to loose stem. There were no findings during this evaluation that indicate the reported devices were defective. No other information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event. There are multiple factors that may contribute to an event that are outside the control of djo surgical. There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event Description
Revision surgery - loosening of ulna stem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDISCOVERY ELBOW
Type of DeviceDISC HUM 4X150MM LT FLANGED C
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, tx 
MDR Report Key10200841
MDR Text Key196765331
Report Number1644408-2020-00492
Device Sequence Number1
Product Code JDC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K013042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial
Report Date 06/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Catalogue Number114914
Device Lot Number920170
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/28/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/19/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 06/26/2020 Patient Sequence Number: 1
-
-