• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number 10605
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/05/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age at time of event: 18 years or older.Device is a combination product.
 
Event or Problem Description
It was reported that stent damage occurred.The target lesion was located in the highly calcified and highly tortuous left circumflex (lcx).A 3.00 x 28 synergy ii stent was unable to cross through a watchdog hemostasis valve and the stent became damaged.The watchdog was removed and replaced with exchanged with another watchdog.It was noted that the stent was damaged by the first watchdog, thus causing stent to deform upon placement in the lcx.The stent was successfully deployed after multiple balloons and the procedure was completed.It was noted that a 2.5 x 20 apex balloon was used to dilate the target lesion.No patient complications were reported in relation to this event and the patient was discharged home the same day.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10200877
Report Number2134265-2020-07988
Device Sequence Number11324358
Product Code NIQ
UDI-Device Identifier08714729840299
UDI-Public08714729840299
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/01/2021
Device Model Number10605
Device Catalogue Number10605
Device Lot Number0024593816
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/05/2020
Initial Report FDA Received Date06/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-