MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC S8 EM ENT; INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC S8 EM ENT; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9735669
Device Problem Imprecision (1307)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/23/2020
Event Type  malfunction  
Manufacturer Narrative
Other relevant device(s) are: product id: 9 735736, serial/lot #: 1.2.0.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system used during a functional endoscopic sinus surgery (fess).It was reported that the surgeon noticed a 5 mm lateral inaccuracy while in the sinuses.This occurred toward the end of the case and the surgeon decided to proceed rather than re-register.The surgeon felt accurate at the start of the case and in the middle.The system would be checked out and logs would be pulled.There was no delay and no impact to the patient.
 
Manufacturer Narrative
H3) the manufacturer representative went to the site to test the navigation system.The system was working as intended and no parts were replaced.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Additional information was received.It was reported that the cause was unknown.
 
Manufacturer Narrative
H2: additional information received.B5 has been updated.H2: additional information received.Section a has been updated.H2: additional information received.Section e has been updated.H2/d10/h3/h6: logs were received for analysis however analysis has not yet been done.Codes 4118, 3233 and 11 are applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
H3) a software analysis was initiated.It was determined that there was insufficient information to determine the cause of the event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Patient archive analysis.It was found that there are many embedded points on patient's right side.Additionally, coverage is minimal.Trace on nose does not go across the bridge of the nose, but rather just a straight line from tip up to the forehead.Green circle of accuracy is minimal and anterior.It was recommended that the surgeon trace across bridge of the nose and to collect more posterior and superior.Also recommended the surgeon try to use a lighter touch when tracing over soft tissue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S8 EM ENT
Type of Device
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key10201003
MDR Text Key198898042
Report Number1723170-2020-01776
Device Sequence Number1
Product Code PGW
UDI-Device Identifier00643169838918
UDI-Public00643169838918
Combination Product (y/n)N
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 09/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735669
Device Catalogue Number9735669
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2020
Initial Date Manufacturer Received 06/23/2020
Initial Date FDA Received06/26/2020
Supplement Dates Manufacturer Received06/24/2020
07/10/2020
07/28/2020
09/16/2020
Supplement Dates FDA Received07/09/2020
07/20/2020
08/07/2020
09/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
"SEE H10...."