| Model Number |
R SERIES |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed an "ecg fault 3" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation; the customer's report was observed during review of the device activity log.However, the customer report was not duplicated with the device after extensive testing.The analog board was replaced as a precaution.The device was recertified successfully and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|