BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 4.0/18; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 364492
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/05/2020
Type of Reportable Event Malfunction
Event or Problem Description
An orsiro drug-eluting sent system was chosen for treatment of a severely calcified lesion (99 percent stenosis degree) in a mildly tortuous lad.Resistance was felt during insertion and the device was removed from the patients body but the stent was missing.Upon an additional angiographic check the dislodged stent was found inside and removed by using a snare.
 
Additional Manufacturer Narrative
The returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation revealed that the returned stent is severely deformed over its entire length.The stent is still attached to the snare which was used for retrieval.The delivery system of the complaint instrument was not returned.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 4.0/18
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key10202632
Report Number1028232-2020-02621
Device Sequence Number11321272
Product Code NIQ
UDI-Device Identifier07640130439078
UDI-Public07640130439078
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/05/2021
Device Model Number364492
Device Catalogue NumberSEE MODEL NO.
Device Lot Number10196193
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2020
Initial Date Received by Manufacturer 06/25/2020
Supplement Date Received by Manufacturer07/31/2020
Initial Report FDA Received Date06/26/2020
Supplement Report FDA Received Date08/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-