• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CORPORATION SABER 4MM4CM 90; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION SABER 4MM4CM 90; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 48004004S
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/03/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the manufacturing documentation associated with lot (82173771) presented no issues during the manufacturing process that can be related to the reported event.Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
As reported, the 4mm x 4cm x 90cm saber balloon catheter was inserted and crossed the lesion and was attempted to be inflated; however, it was already ruptured and could not be inflated.Therefore, device was replaced with a new non-cordis balloon catheter and it could be inflated.The procedure was completed.There was no reported patient injury.The lesion was the right superficial femoral artery.An approach was made from the right common femoral artery.The lesion was chronic total occlusion (cto) and had severe calcification.The device will not be returned for analysis due to the hospital regulation.
 
Additional Manufacturer Narrative
The 4mm x 4cm x 90cm saber balloon catheter was inserted and crossed the lesion and was attempted to be inflated; however, it was already ruptured and could not be inflated.Therefore, device was replaced with a new non-cordis balloon catheter and it could be inflated.The procedure was completed.There was no reported patient injury.The lesion was the right superficial femoral artery.An approach was made from the right common femoral artery.The lesion was a chronic total occlusion (cto) and had severe calcification.The product was not returned for analysis.A product history record (phr) review of lot 82173771 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Vessel characteristics of a chronic total occlusion combined with severe calcification may have contributed to the reported event as calcification is known to cause damage to balloons.According to the warnings in the safety information in the instructions for use ¿prior to angioplasty, the catheter should be examined to verify functionality and integrity, and ensure that its size and shape are suitable for the specific procedure for which it is to be used.Do not use if product damage is suspected or evident.To reduce the potential for vessel damage or the risk of dislodgement of particles it is very important that the inflated diameter of the balloon should approximate the diameter of the vessel just proximal and distal to the lesion.The balloon dimensions are printed on the product label.The compliance table incorporated with the product shows how balloon diameter increases as pressure increases.Do not exceed the rated burst pressure recommended on the label.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Pressure in excess of the rated burst pressure can cause balloon rupture and potential inability to withdraw the catheter through the introducer sheath.Balloon rupture can cause vessel damage and the need for additional intervention.Use only the recommended balloon inflation medium (a 50/50 mixture by volume of contrast medium and normal saline).Never use air or any gaseous medium to inflate the balloon.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event; therefore, no corrective/preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SABER 4MM4CM 90
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60th ave
miami lakes FL 33014
MDR Report Key10203112
Report Number9616099-2020-03765
Device Sequence Number11479427
Product Code LIT
UDI-Device Identifier20705032069851
UDI-Public20705032069851
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K971010
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2022
Device Model Number48004004S
Device Catalogue Number48004004S
Device Lot Number82173771
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2020
Supplement Date Received by Manufacturer07/07/2020
Initial Report FDA Received Date06/26/2020
Supplement Report FDA Received Date07/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
STERLING BALLOON CATHETER.
-
-