| Model Number |
2576-05-029 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
05/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Visual inspection found a small crack in the black marker band and a set screw fell out.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
|
| |
|
Event or Problem Description
|
|
Srom pinnacle hip replacement; the screw in the miller shell fell out during use.No ae to patient.
|
| |
|
Search Alerts/Recalls
|