ABBOTT PROMOTE+ CRT-D W/SJ4 CONNECTOR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3211-36Q
Medical Device Problem Code Over-Sensing (1438)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/11/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
During an in clinic follow-up, post-paced t-wave oversensing was noted on the device.The device was reprogrammed to resolve the event and the patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMOTE+ CRT-D W/SJ4 CONNECTOR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
645 almanor avenue
sunnyvale CA 94085
MDR Report Key10203360
Report Number2938836-2020-07121
Device Sequence Number2526292
Product Code NIK
UDI-Device Identifier05414734503679
UDI-Public05414734503679
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2016
Device Model NumberCD3211-36Q
Device Catalogue NumberCD3211-36Q
Device Lot NumberS000002985
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/11/2020
Initial Report FDA Received Date06/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-