Related manufacturer reference: 3005334138-2020-00235, 3005334138-2020-00261, 3008452825-2020-00329.During a procedure, resistance was noted while trying to advance the brk needle.Force was needed to advance the device.The sheath and needle were replaced, and again the issue was noted.The devices were replaced a third time and resolution was found to complete the procedure.There were no patient consequences.No additional access site was required, however a clinically significant delay of 20-30 minutes was noted.
|
One brk-1¿ transseptal needle, 89 cm length was received for investigation.The needle had been bent.The needle/stylet assembly was inserted into an 8.5f sl1 sheath/dilator assembly from current inventory.Resistance was noted due to the aforementioned damage to the needle.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the received condition of the device and the information provided, the cause of the bend in the needle and subsequent needle insertion difficulty remains unknown.
|