• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. INTERTAN 10S 10MM X 18CM 125D; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. INTERTAN 10S 10MM X 18CM 125D; NAIL, FIXATION, BONE Back to Search Results
Model Number 71675383
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/03/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery upon trying to attach nail to guide, the scrub tech could not get the guide bolt to turn more than an eighth of a rotation.The guide was flush against the top of the nail and the nail was aligned correctly with the axis of the guide, nevertheless after multiple attempts we tried the second bolt in the set with the same results.At this point the dr.Wanted a new nail in fear that if we tried anymore the nail would become cross threaded causing it to bind and presenting issues when trying to remove guide further along in the surgery.A new nail was open and attached to guide with no problems and the surgery was completed as per usual.No delay nor injury reported.
 
Additional Manufacturer Narrative
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The device has signs of damage from attempted use.The medical investigation concluded that no relevant supporting clinical information has been provided to assist with a clinical investigation, and it is unknown if the reported issue caused a delay during the procedure.Therefore based on insufficient information, a thorough clinical assessment cannot be performed at this time.Should any additional clinical information be provided this complaint will be re-evaluated.A dimensional inspection of the device threads found them to be out of specification, confirming the stated failure.This issue was investigated through our internal nonconformance process and was found to be isolated at this time.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have reason to suspect that the product failed to meet product specifications at the time of manufacture.The potential probable cause for this event is likely a manufacturing process error.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERTAN 10S 10MM X 18CM 125D
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10205125
Report Number1020279-2020-02807
Device Sequence Number13980047
Product Code JDS
UDI-Device Identifier00885556105696
UDI-Public00885556105696
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K040212
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71675383
Device Catalogue Number71675383
Device Lot Number20DM05725
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2020
Initial Date Received by Manufacturer 06/03/2020
Supplement Date Received by Manufacturer07/29/2021
Initial Report FDA Received Date06/29/2020
Supplement Report FDA Received Date07/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-