MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. OER-3 100V; ENDOSCOPE REPROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. OER-3 100V; ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-3
Device Problem Failure to Clean Adequately (4048)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The subject device in this report has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the user facility that even though the detergent replacement indicator which indicated the detergent tank became empty blinked, the user reprocessed the unspecified endoscope with the subject device.After the reprocessing, the user used the endoscope to the patient.The field service checked the subject device at the facility and found that there was air in the tube which was connected the detergent tank and the reprocessing basin, consequently the detergent supplying had trouble.There was no report regarding patient injury related to the event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation, therefore omsc could not investigate the subject device.Based on the reported information, omsc concluded the reported phenomenon was attributed to the inappropriate reprocessing by the user, such as the user used the subject device even though the detergent replacement indicator which indicated the detergent tank became empty blinked.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OER-3 100V
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10205152
MDR Text Key213252556
Report Number8010047-2020-03715
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberOER-3
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/04/2020
Initial Date FDA Received06/29/2020
Supplement Dates Manufacturer Received10/20/2020
Supplement Dates FDA Received11/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-