Model Number N/A
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/05/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that debris was identified in the sterile packaging while investigating items in stock.No patients were involved.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Evaluation of the photographs provided confirmed the sterile packaging (blister and foam) is damaged and there is debris inside the sterile packaging which is consistent with the appearance of foam debris from the foam packaging inside the sterile barrier.Therefore, the reported event is confirmed.Sterility was not compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The likely condition of the device when it left zimmer biomet is conforming to specification.The root cause of the reported event it likely to be damage during transit causing the foam packaging to become abraded and shed.The device evaluation found no malfunction and the event did not contribute to injury, therefore this would not be considered a reportable event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional information on reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 MP TYPE1 PPS HO 9.0 MM T1
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10206287
Report Number0001825034-2020-02547
Device Sequence Number14270451
Product Code KWA
UDI-Device Identifier00880304513778
UDI-Public(01)00880304513778
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K110400
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/30/2026
Device Model NumberN/A
Device Catalogue Number51-107090
Device Lot Number3751576
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/05/2020
Supplement Date Received by Manufacturer07/09/2020
Initial Report FDA Received Date06/29/2020
Supplement Report FDA Received Date07/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
-
-