STRYKER GMBH CARBON TUBE, BLUE MONOTUBE TRIAX 20 X 250MM; COMPONENT, TRACTION, INVASIVE Back to Search Results
Model Number 5150-2-411
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Codes Injury (2348); Malunion of Bone (4529)
Date of Event 06/03/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Upon completion of investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
As reported: after application of a blue triax (femur, child) with carbon rod and two 4-pin holders, a pin holder on the rod shifted because it was loosened and the fracture healed with a malposition.Additionally reported: revision surgery necessary with plate osteotomy.
 
Event or Problem Description
As reported: after application of a blue triax (femur, child) with carbon rod and two 4-pin holders, a pin holder on the rod shifted because it was loosened and the fracture healed with a malposition.Additionally reported: revision surgery necessary with plate osteotomy.
 
Additional Manufacturer Narrative
This device is concomitant and did not contribute to the reported failure.The device was actually received fully functional.No loosening was noticed.Therefore, this complaint is closed without further investigation.If any additional information is provided indicating otherwise the record will be reopened and the investigation will be reworked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARBON TUBE, BLUE MONOTUBE TRIAX 20 X 250MM
Common Device Name
COMPONENT, TRACTION, INVASIVE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
MDR Report Key10208486
Report Number0008031020-2020-01804
Device Sequence Number16667901
Product Code JEC
UDI-Device Identifier07613327078978
UDI-Public07613327078978
Combination Product (Y/N)N
PMA/510(K) Number
K963614
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number5150-2-411
Device Catalogue Number51502421
Device Lot NumberBZFXA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2020
Initial Date Received by Manufacturer 06/03/2020
Supplement Date Received by Manufacturer09/24/2020
Initial Report FDA Received Date06/29/2020
Supplement Report FDA Received Date10/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age9 YR
Patient Weight30
-
-