| Model Number |
M00558340 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was used in the esophagus during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2020.According to the complainant, during the procedure, the balloon popped while being inflated to 10mm.Reportedly, the balloon was removed from the patient.The procedure was completed with another cre fixed wire dilatation balloon.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: problem code 1074 captures the reportable event of balloon burst.Block h10: a visual examination of the returned complaint device revealed the balloon was torn longitudinally.This failure is likely due to factors or conditions related to the procedure that could have affected its performance and its intended purpose such as lesion characteristics, handling of the device, the technique used by the physician, and/or normal procedure difficulties.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was used in the esophagus during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6), 2020.According to the complainant, during the procedure, the balloon popped while being inflated to 10mm.Reportedly, the balloon was removed from the patient.The procedure was completed with another cre fixed wire dilatation balloon.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be fine.
|
| |
|