Model Number DVAB1D1
Medical Device Problem Codes High impedance (1291); Mechanical Problem (1384); Misconnection (1399); Over-Sensing (1438); High Capture Threshold (3266)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that within a few days of a implantable cardioverter defibrillator (icd) change out procedure where it was indicated that there had been difficulty fully inserting the pace/sense pin, an alert was triggered due to high/undefined pacing impedance.Oversensing was noted on stored episodes and the sensing integrity counter (sic) had incremented.High pacing threshold was also noted.Reprogramming was performed.The right ventricular (rv) lead and icd remain in use.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the impedance of the right ventricular pacing lead was beyond the expected upper range.Analysis of the device memory indicated pacing capture threshold in the right ventricle was elevated.Analysis of the device memory indicated oversensing due to non-physiologic signals/sensing integrity counter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
It was further reported that there was a setscrew issue with the device and the lead was not fully deployed.A revision procedure took place.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIA AF XT VR
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key10209548
Report Number3004209178-2020-11264
Device Sequence Number2107612
Product Code LWS
UDI-Device Identifier00643169720619
UDI-Public00643169720619
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P980016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/14/2020
Device Model NumberDVAB1D1
Device Catalogue NumberDVAB1D1
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/15/2020
Supplement Date Received by Manufacturer06/30/2020
07/07/2020
Initial Report FDA Received Date06/29/2020
Supplement Report FDA Received Date07/07/2020
07/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6944-58 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age69 YR
Patient Weight99
-
-