| Model Number |
1488210105 |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported there was loss of image for a minute or less during the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: trappers on site stopped.The failure(s) identified in the investigation is consistent with the complaint record.Root cause: cable issue, bad grounding.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported there was loss of image for a minute or less during the procedure.
|
| |
|