• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MITG - RF SURGICAL SYSTEMS SITUATE; GAUZE/SPONGE, INTERNAL, X-RAY DETECTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MITG - RF SURGICAL SYSTEMS SITUATE; GAUZE/SPONGE, INTERNAL, X-RAY DETECTABLE Back to Search Results
Model Number G0804-12P02C-1
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/08/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic conducted an investigation based upon all information received.The involved device was not returned.Visual inspection identified that the cotton was fraying.Loose strings were shown in the picture.The gauze was removed from the package and was unfolded.It was reported that the gauze showed linting.A related device issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are also thoroughly reviewed prior to release to ensure that products meet all medtronic quality specifications.The instructions included with this device provide the following guidance: the gauze sponges are made with natural cotton and may produce lint or fibers when unfolded or manipulated.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, prior to use, when they opened the pack, they noticed that the gauze was falling apart and unraveling.A picture was submitted which showed that the gauze's cotton strands were separating.There was no patient involvement.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found that the opened package had excessive fabric detached from the gauze.It was reported that the product was linting.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are also thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the gauze sponges are made with natural cotton and may produce lint or fibers when unfolded or manipulated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITUATE
Common Device Name
GAUZE/SPONGE, INTERNAL, X-RAY DETECTABLE
Manufacturer (Section D)
MITG - RF SURGICAL SYSTEMS
2101 faraday
carlsbad CA 92008
MDR Report Key10210460
Report Number3005883396-2020-00031
Device Sequence Number2565238
Product Code GDY
UDI-Device Identifier10884521700833
UDI-Public10884521700833
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG0804-12P02C-1
Device Catalogue NumberG0804-12P02C-1
Device Lot Number200105AT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/31/2020
Initial Date Received by Manufacturer 06/08/2020
Supplement Date Received by Manufacturer07/29/2020
Initial Report FDA Received Date06/30/2020
Supplement Report FDA Received Date08/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-