| Model Number |
U4150330 |
| Date of Event |
06/09/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.Expiry date (12/2022).
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure through tibial artery, the pta balloon allegedly had a pinhole ruptured at 4 atm in the posterior tibial (pt) artery.It was further reported that the treatment plan changed to at/pedal loop.The procedure was completed using another device.The patient status is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: a lot history review was conducted and it was determined that a device history record review was not required.Investigation summary: the investigation is inconclusive for the reported balloon rupture, as the device was not returned for evaluation.The definitive root cause for the reported balloon rupture could not be determined based upon available information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(expiry date: 12/2022).
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure through tibial artery, the pta balloon allegedly had a pinhole ruptured at 4 atm in the posterior tibial (pt) artery.It was further reported that the treatment plan changed to at/pedal loop.The procedure was completed using another device.The patient status is unknown.
|
| |
|