| Model Number |
MMT-723NAH |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
06/24/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they experienced high blood glucose.The customer¿s blood glucose level was 430 mg/dl and 267 mg/dl which was treated with manual injection.Customer stated that drive support cap was normal.Customer had been using insulin pump system within 48 hours of reported high blood glucose event.Customer was neither in emergency room nor admitted into hospital.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test, rewind, basic occlusion, occlusion, prime or a33 and excessive no delivery test.(b)(4).
|
| |
|