• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) ANTHEM RF CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) ANTHEM RF CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3210
Medical Device Problem Codes Failure to Interrogate (1332); Wireless Communication Problem (3283)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); Insufficient Information (4580)
Date of Event 06/17/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
It was reported that patient¿s device was not able to be interrogated.The interrogation was attempted using inductive telemetry, different programmer, clearing the environment to remove effect of electromagnetic interference (emi), magnet, radio frequency telemetry test.All troubleshooting failed.The decision was made to interrogate the device in a different room.If interrogation fails, the device replacement was discussed.Further information was requested but was not received.
 
Event or Problem Description
New information received notes that the device was explanted and replaced.Elective replacement indicator (eri) was also noted.
 
Additional Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Additional Manufacturer Narrative
The reported field event of failure to interrogate was not confirmed in the laboratory.The device was tested on the bench, and no anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHEM RF CRT-P
Common Device Name
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
MDR Report Key10211387
Report Number2938836-2020-07174
Device Sequence Number2120841
Product Code NKE
UDI-Device Identifier05414734503785
UDI-Public05414734503785
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P030035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2014
Device Model NumberPM3210
Device Catalogue NumberPM3210
Device Lot Number3782427
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/17/2020
Supplement Date Received by Manufacturer11/12/2020
02/18/2021
Initial Report FDA Received Date06/30/2020
Supplement Report FDA Received Date12/05/2020
02/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-