| Model Number |
PM3210 |
| Medical Device Problem Codes |
Failure to Interrogate (1332); Wireless Communication Problem (3283)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); Insufficient Information (4580)
|
| Date of Event |
06/17/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
It was reported that patient¿s device was not able to be interrogated.The interrogation was attempted using inductive telemetry, different programmer, clearing the environment to remove effect of electromagnetic interference (emi), magnet, radio frequency telemetry test.All troubleshooting failed.The decision was made to interrogate the device in a different room.If interrogation fails, the device replacement was discussed.Further information was requested but was not received.
|
| |
|
Event or Problem Description
|
|
New information received notes that the device was explanted and replaced.Elective replacement indicator (eri) was also noted.
|
| |
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported field event of failure to interrogate was not confirmed in the laboratory.The device was tested on the bench, and no anomalies were found.
|
| |
|